Show simple item record

dc.contributor.authorMuhamad Khairul, Ali Hassan
dc.contributor.authorNagamune, Kouki
dc.date.accessioned2014-04-29T06:55:44Z
dc.date.available2014-04-29T06:55:44Z
dc.date.issued2013-07
dc.identifier.citationp. 7180-7183en_US
dc.identifier.isbn978-145770216-7
dc.identifier.issn1557-170X
dc.identifier.urihttp://ieeexplore.ieee.org/xpl/articleDetails.jsp?arnumber=6611214
dc.identifier.urihttp://dspace.unimap.edu.my:80/dspace/handle/123456789/34173
dc.descriptionProceeding of The Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2013 (EMBC 2013) at Osaka, Japan on 3 July 2013 through 7 July 2013.en_US
dc.description.abstractPedicle Screw (PS) was originally accomplished for Spinal fixation but it has several limitations. If the pedicle screw passes though the cortex bone in surgery, it has a risk to damage the spinal cord and vertebral artery, which can be caused to a serious problem such as paralysis. Therefore, it should be avoided by all possible means. In current situation, it depends on a palpation of doctor to judge the boundary between the cortex and cancellous bone. Although many instrumentation has been described in several clinical studies, there are still lack of data in the literature concerning the measurement of bone thickness in real time mode. Most of the measurements of the bone thicknesses were based on CT Scan machine which is off-line technique. Therefore, the purpose of this prospective study was to develop a real time measurement of bone thickness for safety purpose of pedicle screw insertion. A total of 12 data was collected in each experiment. Ultrasound echo signal for each specimen was measured and used to measured bone thickness. Then, the results were compared with manual measurement of bone thickness which is by using a ruler. The percentage different of bone thickness was small for both methods which were 8.86 % for first method and 15.1 % for second method. This measurement values showed that the accuracy of bone thickness more than 84 % for both method. As a conclusion, both methods were suitable to use as a bone thickness measurement technique for pedicle screw insertion application.en_US
dc.language.isoenen_US
dc.publisherInstitute of Electrical and Electronics Engineers (IEEE)en_US
dc.relation.ispartofseriesProceeding of The Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2013 (EMBC 2013);
dc.subjectPedicle screw insertionen_US
dc.subjectThickness measurementen_US
dc.subjectBoneen_US
dc.titleA new technique of bone thickness measurement for Pedicle Screw insertion safety purposeen_US
dc.typeWorking Paperen_US
dc.contributor.urlkhairulhassan@unimap.edu.myen_US
dc.contributor.urlnagamune@me.his.u-fukui.ac.jpen_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record